Bench-Top Analyzer
The SINDIE-7039 Bench-Top analyzer is a compact, easy to use sulfur analyzer
designed for petroleum fuels from ultra-low sulfur diesel and gasoline to heavy
fuel oil. The SINDIE-7039 delivers unprecedented accuracy and precision,
and provides the ultimate solution for the petroleum industry where
reliability and speed are critical. Breakthrough Monochromatic
WD XRF technology offers a limit of detection of 0.4 ppm
wt. with a dynamic range up to 3000 ppm wt. This
direct measurement technique does not require
sample conversion, consumable gases or high
temperature operation.
The SINDIE-7039 design is ideal for laboratory
and industrial use and requires minimum
maintenance.

Application Areas

Orange DotTotal sulfur analysis from ULS fuels up
white dotto 3000 ppm
odotFrom gasoline to heavy fuel oil
odotFor pipeline terminals, refineries and
white dottest laboratories

Features and Benefits

odotLOD: 0.4 ppm wt.
white dotDynamic range: 0.4-3000 ppm wt.
odotRepeatability:

S Concentration Std. Dev.
1 ppm 0.1 ppm
10 ppm 0.4 ppm
100 ppm 1.0 oom
500 ppm 1.7 ppm

Orange DotFits on any bench in any lab:
White Dot14.5"(W) x 19.5"(D) x 13.5"(H)
Orange DotPlug and go design: standard wall power is only utility required
Orange DotUser-friendly with touch screen interface
Orange DotFlip-lid design for easy sample cup introduction
Orange DotMeasurment time 30s - 300s programmable
Orange DotOne calibration for diesel and gasoline matrixes over full dynamic range
Orange DotExtremely low maintenance:
White DotNo conversion gases
White DotNo columns
White DotNo heating elements
White DotNo quartz tubing
Orange DotField replaceable air-cooled excitation tube
Orange DotNo sample conversion or combustible gases required
Orange DotPolyamide x-ray windo, no exposed Be windows

Typical SINDIE-7039 Configuration

Sindie Data 1

Figure 1
Analytic Engine Configuration

Monochromatic excitation=Extremely low background

No moving parts in analyzer engine

No consumables or sample conversion required

Sindie Data 2Figure 2

Linearity 0-500 ppm

R2 = 0.9999


Orange DotTable 1. Repeatability
Successive sulfur measurements of diesel specimen
at various sulfur concentrations on same analyzer.
measurement time:300 seconds.
Sindie Data 3
Orange DotTable 2. Reproducibility Values at 95%
Confidence Level.
An interlaboratory study among six labs, using six different analyzers, analyzed 10 diesel fuels and 9 gasolines (some containing oxygenates). Samples were analyzed in duplicate and back-to-back. None of the labs used matrix correction calculation For information, contact your XOS representative.
Sindie Data 4
Test Method: In accordance with
ASTM Standard Test Method D-7039
-------------------
Dimensions:
14.5"(W) x 19.58"(D) x 13.5"(H)
----------
Power:
100-120 VAC, 47-63 Hz at 6.0 Amps
200-240 VAC, 47-63 Hz at 6.0 Amps
----------
Other Utilities:
None
----------
Sample Introduction:
Maximum sample cup volume: 15cc
I/O Ports: Ethernet 10/100 base T or RS232
Optional Computer: Pentium, 100 Mhz, 32 MB RAM
----------
Interface:
Windows 98 or newer operating system
----------
Ambient Temperature Requirements:
5-40 C (40-10 F)
----------
Dynamic Range:
0.4 ppm - 3000 ppm (wt.)
----------
Measurement Time:
30-300 seconds